Fabrication of Simple Imaging Chamber
1. Prepare the Millicell Standing Cell Culture Insert by removing the bottom legs with a blade (see Note 5).
2. Siliconize the glass bottom Petri dish (see Note 6).
3. Your device is now ready for sample loading.
Fabrication of Advanced Imaging Chamber
1. If creating a new design, start with hand drawings.
2. Once drawn, create the design on the computer (any software can be used) and surround it with a border (see Note 7).
3. Clip a screenshot of the design.
4. Open the Silhouette software.
5. Paste the design into Silhouette.
6. Open the trace panel and trace the design (see Figure 1A).
7. To help size the design, use the grided view (Click on view à show grid).
8. Each square is 1 inch so change the design size by clicking and dragging points.
9. Measure channel width with the line drawing tool (see Figure 1B) (see Note 8).
10. The overall size should be somewhat smaller than a 24 x 60 mm cover glass as this will be the base of the device (see Note 9).
11. Create circles between 1.5 and 2.0 mm in diameter and place them at each entry and exit point.
12. Copy the entire design and paste it nearby.
13. Remove all channels leaving only the entry and exit circles (this will be the top layer of the design).
14. Box your design so that it will be cut out (see example in Figure 2A).
15. When placing your design in the grid, Do Not place it in the top left corner as the Cameo cutter tests its readiness here.
Cutting Design
16. With gloves (to prevent oil transfer from hands), replace the protective cover on the cutting mat with a piece of laminate with the shiny side of the laminate face down (see Note 10).
17. Using a roller, remove any air bubbles from underneath the laminate, fully securing it to the cutting mat.
18. Line the cutting mat up with the blue (line) guide on the CAMEO and hit “Load cut mat”
19. In your design file, click Send.
20. In this menu set the force, speed, material, passes, and blade depth (see Note 11).
21. Connect your computer to the Cameo cutter (see Note 12).
22. Remove anything from the back of the machine as it will use this space to pull the cutting mat through.
23. Make sure the design that you have created in silhouette will be cut on an area where you have placed laminate.
24. The Cameo cutter should indicate it’s “Ready” and “TEST” should be selectable.
25. Test the readiness by selecting “TEST”.
26. Click Send in the Sendwindow.
27. Once complete hit Unload.
Laminating Design
28. Turn on the laminator and allow it time to warm up (see Note 13).
29. Remove the cut design from the laminate using a pair of tweezers.
30. With small pieces of double-sided tape, assemble the design in the proper layered order with the coverslip on the bottom (see Figure 2B).
31. Make sure to have all adhesive, matte, sides facing inwards, away from the laminator (see Note 14).
32. Once assembled, run the cut design attached to the coverslip through the laminator (see Note 15).
33. Perforate furniture bumpers (see Note 16).
a. Punch holes in larger bumpers with the revolving punch pliers.
b. Using biopsy punches, create holes in the smaller bumpers.
34. Remove cut material.
35. Place the perforated furniture bumpers over the exit and entry holes (see Note 17).
36. Your device is now complete and ready for sample loading (see Figure 2B).
37. Assemble tubing and syringe with the Polyetheretherketone (PEEK) tubing adapter.
38. Then begin by cleaning the tubing that will be used to supply media to the microfluidic device by flushing the tubing with 70% Isopropyl Alcohol (IPA) and air, alternating between them.
39. Do this three times ending with air and adding solutions to the syringe using a 1 mL pipette by removing the plunger of the syringe.
40. Rince a final time with PBS or Grace’s media.
41. Place aside and retrieve for final assembly.
Dissection of Drosophila Tissue:
1. Obtain wandering larvae that contain a genetic background with a calcium biosensor such as a form of GCaMP (see Note 18).
2. Prepare the dissection microscope by shifting the working area to the black surface as this will help make the larva more visible and dissection easier.
3. Place a siliconized black glass dissection plate over the microscope stage.
4. Prepare four small Petri dishes with 2 containing PBS, one with 70% isopropyl alcohol (IPA), and one with a small amount of deionized water (DI) (see Note 19).
5. Retrieve larvae using blunt forceps and rinse in first dish of PBS, making sure to remove any debris.
6. Transfer to 70% IPA and let sit for ~3 minutes, swirling occasionally.
7. After sterilizing in 70% IPA, rinse in second dish of PBS.
8. Then place larvae in the final dish with DI water until ready to dissect (see Note 20).
9. Many larvae can be “stored” like this at a time so repeat steps 5-8 as many times as needed.
10. Clean the dissection plate with a Kimwipe and 70% IPA.
11. Place a drop of Grace’s Medium (~30μL) onto the glass plate (see Note 21).
12. Using the blunt tweezers, place a larva in the drop.
13. Under the microscope, locate the posterior and using the blunt tweezers grab the larva around the middle (closer to the posterior) (see Note 22).
14. With the sharp tweezers, grab the larva next to the blunt ones and tear the posterior part away. The posterior half can be discarded (into an adjacent “garbage” drop, for example).
15. Now grab the torn end with the blunt tweezers to stabilize the larva so that the sharp tweezers can be inserted into the anterior end (see Note 23).
16. With the blunt tweezers, invert the larva by pushing it onto the sharp tweezers.
17. This will expose the larva tissue which contains the imaginal discs.
18. Remove the sharp tweezers after the larva has been inverted.
19. To locate the discs, begin to remove some of the tissue like the gut and fat.
20. After some tissue removal, locate desired tissues like the developing wing imaginal discs
21. These can be found by first finding the trachea as this is a bright white line of tissue that runs along the entire larva, is easy to identify, and has the wing imaginal discs attached to it at the anterior end (along with the thoracic leg and haltere discs).
22. If targeting the brain, it is found at the anterior end of the larvae, surrounded by additional developing tissue like the eye discs.
23. When desired tissue is identified, isolate it while paying attention not to stretch, scratch, or tear it.
24. Once isolated, wash tissues by transferring them through a series of drops using a 2 μL Micropipette.
25. Tissues can be stored in a drop of medium until all dissections are complete (see Note 24).
Preparation of Imaging Chambers for Sample Loading
Simple Imaging Chamber
1. Add 20 μL of Grace’s medium to the center of the glass bottom Petri dish.
2. Transfer dissected and isolated tissues to the drop of medium using a 2 μL Micropipette.
3. Orient the tissue as required for the microscope you will be imaging with (see Note 25).
4. Place the prepped Millicell Standing Cell Culture Insert (phantom legs down) over the tissue, paying attention to not use lateral movements, which can disrupt the orientation of tissue (see Note 26).
5. Add 100-150 μL of Grace’s medium to the top of the Millicell Standing Cell Culture Insert.
6. Place enough (100-150 μL) embryo oil over the Grace’s medium to completely cover it.
7. Place a rolled-up Kimwipe around the inner walls of the Petri dish and wet it with 400 μL of PBS or surround the outside of the imaging chamber with Embryo oil (100-150 μL) (see Note 27).
8. You are ready to transfer the chamber to the microscope (see Note 28).
Advanced Microfluidic Imaging Chamber
1. Retrieve a microfluidic device (see Note 29).
2. Partially fill the larger exit channel (large bumper) with Grace’s media.
3. Place the desired tissue in the larger opening using a 2 μL Micropipette.
4. Use a wire to push the tissue down and slightly into the channel, orienting it properly (see Note 25).
5. With a 20 μL Micropipette, slowly suck the wing disc down into the channel by pipetting volume up from the entrance channel (small bumper) until it has reached a desired location in the design (see Note 30).
6. Once the microfluidic device has been prepared, retrieve the cleaned tubing
7. With a new syringe (and attached needle), suck up ~1 mL of Grace’s media and remove as many air bubbles as you can in the syringe by taping the tip and expelling some media.
8. Remove the needle and replace the syringe that is attached to the tubing with this new one loaded with media (see Note 31).
9. Wash about 0.1 – 0.2 mL of media through the tubing, ensuring all air is removed.
10. With the syringe and microfluidic device prepared, mount the syringe into the syringe pump, which should be placed next to the microscope.
11. Using the menu on the syringe pump, push out a small drop of media and then insert the tubing into the entrance of the entrance channel (small bumper) on the microfluidic device.
12. Set the flow rate on the syringe-pump (see Note 32).
13. Place the imaging chamber onto the microscope and begin imaging.
Live Calcium Imaging of Drosophila Tissue
It is recommended to utilize a spinning disc confocal microscope for improved acquisition speed with conserved resolution. The chosen interval and light exposure levels are essential for proper image acquisition and downstream analysis. This will largely depend on the signaling pathway that will be observed and biosensors being used (see Note 18). Generally, for long-term imaging (8+ hours) intervals of 5-6 min or larger are recommended to avoid damage to cellular tissue. Shorter intervals of 10-30 seconds can be used but will limit the length of imaging to about 2 hours before the tissue’s health starts to become impaired. Additionally, laser power is recommended to be maintained at the lowest detectable setting as is allowed by the system. This will also help prevent tissue damage.
It is also important to understand that the working distance of spinning confocal microscopes are limited to boost resolution. Therefore, the imaging plan chosen is significant when acquiring data. Either a single z-slice should be imaged, or small z-stacks taken. Larger z-stacks can be acquired with larger step sizes; however, this is not recommended as the resolution of the z-axis will be severely limited. The imaging plan should be chosen based on the dynamics to be observed and the tissue oriented appropriately (see Note 25).
Calcium Analysis
Following the imaging of calcium dynamics, is a workflow of image analysis around the acquired fluorescent signals. This usually involves denoising, motion correction if drift is present, segmentation and identification of regions of interest (ROI) (cells experiencing calcium spikes), and signal extraction. This is then ultimately followed by the analysis of the calcium transients and their interpretation. Many platforms have been created for this over a variety of cell types19–24. Here we utilize and demonstrate analysis with CaImAn—an open-source library that can be used in either MATLAB or Python for calcium image analysis19. We have adapted here for use in the developing wing imaginal disc instead of its original use in neurons (Figure 3). This tool takes time-lapse videos of fluorescent capture (Figure 3A) and begins by correcting for motion, which is followed by ROI identification (Figure 3B), and finally data extraction (Figure 3C).
Modifications will likely be required and depends on the platform chosen. Key parameters that must be accounted for in the analysis include the interval rate of capture and cell size, along with the dynamics of the calcium indicator. When selecting an image analysis method or platform, one must also factor into account the amount of prior coding experience needed. Many of the platforms, like CaImAn, allow for the specification of the system to sample needs but can require an advanced understanding of computer languages. Others, like EZcalcium, include a graphical user interface and can be navigated by those with little coding experience20. In conclusion, the post-acquisition analysis workflow involves several decisions to quantify and interpret the results. Several open-access image analysis methods are available. However, regardless of the chosen platform, customization and understanding of coding languages may be necessary, although user-friendly options are available. Thus, selecting the appropriate platform requires consideration of specific experimental needs and the level of coding expertise available.