See Figure 1 for schematic representation
Sample collection and storage:
1. Blood was collected with anti-coagulant for plasma and centrifuged human peripheral blood at 1500 x g for 10 min at room temperature
2. Collect plasma
3. Store at -80ºC until analysis
Protein precipitation:
4. Allow the sample to gently thaw on ice
5. Prepare methanol containing internal standard (methanol-IS; 500pg in 2mL of methanol/0.5mL of plasma) to facilitate quantification (d4-PGE2, d5-RvD2, d5-LXA4, d4-LTB4, d8-5S-HETE, d5-LTC4, d5-LTD4 and d5-LTE4)
6. Place methanol-IS on ice for 45 min
7. Put 1mL of plasma in a 15 mL falcon tube
8. Add 4 mL of cold methanol-IS
9. Vortex for 10 sec
10. Keep samples at -20°C for a minimum of 45 min to allow for protein precipitation
11. Samples can either be stored -80ºC or processed for lipid mediator extraction using solid phase extraction techniques
Solid Phase Extraction:
12. Centrifuge samples at 2000 x g for 10 min at 4°C
13. Transfer 2.5 mL of supernatant into round bottom borosilicate tubes
14. Store any remaining samples at -80ºC
15. Place samples in TurboVap LV setting the water bath to 37°C under a gentle flow of Nitrogen gas
16. Maintain samples in the evaporator until the final volume of methanol is less the 1mL
17. Place samples on ice
18. Critical: Samples should not be stored and need to be processed ASAP
19. Turn on and set-up Extrahera
a. Add new Solid-phase ISOLUTE C18 500 mg/3 mL columns to the rack
b. Add new collection tubes
c. Fill the solvent stock bottles
d. Prime the different solvent feeders
e. Turn on the vacuum pump (Adjust the pressure in accordance with the number of samples you plan to run: 3 bar = ¼ , 4 bar = ½ and 6 bars = full plate)
f. Turn on the nitrogen supply (Adjust the pressure in accordance with the number of samples you plan to run: 6 bar = ¼ , 8 bar = ½ and 10 bars = full plate)
g. Empty the waste
h. Fill the tips
20. Load the samples on Extrahera and start the run
a. Wash columns with 3 mL methanol
b. Condition columns with 3 mL of water
c. Acidified samples by adding pH 3.5 water (HCl) to a final volume of 10 mL
d. Load samples onto the C18 SPE columns
e. Wash columns with 2 mL of water to remove the acid
f. Wash columns with 6 mL hexane to remove complex lipids
g. Elute mediators into collection tubes using 5 mL methyl formate
h. Elute conjugated mediators into separate collection tubes using 5 mL methanol
21. Transfer the collected fractions to 15 mL falcon tubes
22. Rinse the collection tubes with 1.5 mL methanol and pool with the respective fraction
Sample preparation for LC-MS/MS
23. Place samples in TurboVap LV setting the water bath to 37°C under a gentle flow of Nitrogen gas
24. Place an equal volume to the extracted sample of methanol-IS into a separate 15 mL falcon tube and place it in evaporator with samples. This will be used as a reference to calculate sample recovery
25. Check regularly the solvent levels
26. When ~95% of the solvent has evaporated, rinse the wall with 2 mL methyl formate
27. Check regularly the solvent levels
28. When ~95% is evaporated, rinse the wall with 2 mL methanol
29. Leave samples in TurboVap LV until all the solvent has evaporated
30. Add 40 μL of methanol/water (1:1, vol/vol)
31. Vortex 10 sec
32. Immediately centrifuge samples at 2000 x g for 2 min at 4°C
33. Transfer the supernatant into borosilicate vial inserts placed in a 1.5 mL Eppendorf tubes
34. Centrifuge at 10000 x g for 10 sec at 4°C
35. Check the vial insert for the presence of pellet
a. If no pellet is visible, transfer the insert into an injection tube
b. If a pellet is visible, repeat step 33-34-35
36. Place the samples in the automated injector ASAP
Instrument step for LC-MS/MS – Chromatography Setup
37. Start the LC-MS/MS system
Methyl formate fraction profiling instrument setup
38. An Agilent Poroshell 120 EC-C18 column (100 mm x 4.6 mm x 2.7 µm) is kept at 50°C
39. The initial mobile phase consists of methanol/water/acetic acid 20:80:0.01 (vol/vol/vol)
40. The flow rate is maintained at 0.5 mL/min
41. The mobile phase is ramped to 50:50:0.01 (vol/vol/vol) over 0.5 min
42. The mobile phase is ramped to 80:20:0.01 (vol/vol/vol) from 2 min to 11 min
43. The mobile phase is maintained till 14.5 min
44. The mobile phase is ramped to 98:2:0.01 (vol/vol/vol) for the next 0.1 min
45. The mobile phase is maintained at 98:2:0.01 (vol/vol/vol) for 5.4 min
46. Inject 35 µL of the sample
47. Include in the sequence reference standards and the 100% IS (40 µL injection)
Methanol fraction profiling instrument setup
48. An Agilent Poroshell 120 EC-C18 column (100 mm x 4.6 mm x 2.7 µm) is kept at 50°C
49. The initial mobile phase consists of methanol/water/acetic acid of 20:80:0.5 (vol/vol/vol)
50. The flow rate is maintained at 0.6 mL/min
51. The mobile phase is ramped to 55:45:0.5 (vol/vol/vol) over 0.2 min
52. The mobile phase is maintained for 1 min
53. The mobile phase is ramped to 70:30:0.5 (vol/vol/vol) over 5min
54. The mobile phase is ramped to 80:20:0.5 (vol/vol/vol) for 2 min
55. The mobile phase is maintained for 3 min
56. The mobile phase is ramped to 98:2:0.5 (vol/vol/vol) over 3 min
57. The mobile phase is maintained for 2 min
58. Inject 35 µL of the sample
59. Include in the sequence reference standards and the 100% IS (40 µL injection)
Chiral LC-MS/MS
60. A Chiralpak AD-RH column (150 mm × 2.1 mm × 5 μm) is kept at RT
61. The mobile phase consists of methanol/water/acetic acid 95:5:0.01 (v/v/v)
62. The flow rate is maintained at 0.15 mL/min
63. The mobile phase is maintained for 15 min
64. Inject no more than 10 µL of sample
65. Include in the sequence reference standards and the 100% IS
Instrument step for LC-MS/MS – Mass spectrometry Setup
Multiple reaction monitoring (MRM) method setup for mediators eluted in the methyl formate fraction
66. QTrap 5500 or QTrap 6500+ are operated in negative mode
67. Source parameters (probe position, electrode position, curtain gas, collisionally activated dissociation gas, electrode voltage, source temperature, and source gas) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 1 and 2 for parameters used)
68. Individual MRM transitions (Q1 and Q3) and respective parameters (declustering potential, entrance potential, collision energy and collision cell exit potential) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 3 and 4 for parameters used)
69. Select Scheduled MRM option
70. Set data acquisition window to 90 seconds
MRM for method setup for mediators eluted in the methanol fraction
71. QTrap 5500 or QTrap 6500+ are operated in positive mode
72. Source parameters (probe position, electrode position, curtain gas, collisionally activated dissociation gas, electrode voltage, source temperature, and source gas) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 5 and 6 for parameters used)
73. Individual MRM transitions (Q1 and Q3) and respective parameters (declustering potential, entrance potential, collision energy and collision cell exit potential) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 3 and 4 for parameters used)
MRM for chiral analysis:
74. QTrap 5500 or QTrap 6500+ are operated in negative mode
75. Source parameters (probe position, electrode position, curtain gas, collisionally activated dissociation gas, electrode voltage, source temperature, and source gas) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 1 and 2 for parameters used)
76. Individual MRM transitions (Q1 and Q3) and respective parameters (declustering potential, entrance potential, collision energy and collision cell exit potential) are determined using manual tune tool from analyst software by T-infusion of synthetic or authentic standards in phase (see Table 3 and 4 for parameters used)
Information Dependent Acquisition (IDA)-triggered Enhanced Product Ion (EPI) Scan
77. To collect MS/MS spectral information from the same samples an IDA experiment is to be employed
78. In the IDA Ion exclusion tab list the parent ions of the internal standards that are being used for the analysis to minimize interference from these molecules in data collection
79. Set the polarity of the assay in the EPI scan window as appropriate
80. Set the m/z window for ion detection in the EPI window to 100-400 for negative ion mode experiments and 100-680 for positive ion mode experiments
Data analysis:
81. The entire analysis is conducted using Analyst software 1.6.3
82. Build a quantitation method using your standard reference mix
83. Load your samples using the created method
84. Integrate the internal standards noting the retention times of each internal standard in your samples and standards
85. Identify each LM by matching its retention time to synthetic or authentic standards with maximum drift between the expected retention time and the observed retention time window of 0.05 seconds
86. Each internal standard represents specific region of the chromatographic profile. Use:
a. d4-PGE2 for AA-derived prostaglandins and thromboxane
b. d5-RvD2 and d5-LXA4 for trihydroxylated species DHA/DPA and AA/EPA-derived respectively
c. d4-LTB4 for dihydroxylated species DHA/DPA/AA/EPA-derived
d. d8-5S-HETE for monohydroxylated species DHA/DPA/AA/EPA-derived
e. d5-LTC4 for glutathionyl conjugated species DHA/AA-derived
f. d5-LTD4 for cysteinylglycinyl conjugated species DHA/AA-derived
g. d5-LTE4 for cysteinyl conjugated species DHA/AA-derived
87. Internal standards are used to:
a. calculate the expected RT for the analytes of interest in each sample by comparing the relative retention time of the analyte with the respective internal standard in the standard mix.
b. calculate recoveries, comparing the amount of each of the internal standards present in the samples with the amount added, as calculated from the 100% sample.
88. Verify that each peak has at least 4 data points and > 2000 counts
89. Integrate each LM
90. Using the Analyst Explore mode extract the EPI data and for each dataset, for each mediator, in a minimum of one sample per dataset and within the same region of the chromatogram where the peak is observed, match the MS/MS spectrum to that of a reference standard. For a positive match the MS/MS spectrum needs to contain a minimum of 6 diagnostic ions that match those assigned within the reference MS/MS spectrum, with one of these being a backbone fragment.
91. Proceed to quantitation using calibration curves obtained for each mediator construed with the following mediator concentrations: 0 (blank) 0.78, 1.56, 3.12, 6.25, 12.5, 25, 50, 100 and 200 pg.
92. Surrogate molecules carrying similar physical properties are used for molecules where standards are not available